MEASUREMENT OF ENDOTRACHEAL TUBE CUFF PRESSURE IN MECHANICALLY-VENTILATED PATIENTS ON ARRIVAL TO INTENSIVE CARE UNIT - A CROSS-SECTIONAL STUDY

Abstract

Arun Kumar Ajjappa

BACKGROUND
The monitoring of Endotracheal Tube (ETT) cuff pressure in intubated patients on arrival to intensive care unit is very essential. The cuff pressure must be within an optimal range of 20-30cm H2O ensuring ventilation with no complications related to cuff overinflation and underinflation. This can be measured with a cuff pressure manometer.
The aim of the study is to measure the endotracheal tube cuff pressure in patients on arrival to intensive care unit and to identify prevalence of endotracheal cuff underinflation and overinflation.
MATERIALS AND METHODS
A cross-sectional study was done on mechanically-ventilated patients who were intubated in casualty (emergency department) on arrival to intensive care unit in S.S. Institute of Medical Sciences and Research Centre, Davangere. About 50 critically-ill patients intubated with a high volume, low pressure endotracheal tube were included in the study. An analogue manometer was used to measure the endotracheal tube cuff pressure. It was compared with the recommended level. The settings of mechanical ventilation, endotracheal tube size and peak airway pressure were recorded.
RESULTS
It was found that the mean cuff pressure was 64.10 cm of H2O with a standard deviation of 32.049. Of the measured cuff pressures, only 2% had pressures within an optimal range (20-30cm of H2O). 88% had cuff pressures more than 30cm of H2O. The mean peak airway pressure found to be 20.50cm of H2O with a Standard Deviation (SD) of 5.064.
CONCLUSION
This study is done to emphasise the importance of cuff pressure measurement in all mechanically-ventilated patients as cuff pressure is found to be high in most of the patients admitted to intensive care unit. Complications of overinflation and underinflation can only be prevented if the acceptable cuff pressures are achieved.